Structural Biology Center at ANL
ANL :: SBC :: Facilities :: Limits
  19 ID & 19 BM
  Biochemistry Lab
  Software & Hardware
  Laboratory and Office Module
 
   About Us


   SBC Publications



   Staff



   Events




Search


Search WWW
sbc.anl.gov

 

.................................................



 

Limits

Limits for 19-BM


Parameter

Limits

Photon Energy (keV)

 

lower limit

6.0

upper limit

13.5

Energy Bandwidth

 

ΔE/E @ 13.5 keV

2.5 x 10-4

Minimum Focus Achieved (mm)

 

vertical

0.070

horizontal

0.070

Maximum Accepted Divergence from Source (mrad)

 

vertical @ 12.0 keV

0.03

horizontal @ 12.0 keV

0.55

Beam Position Stability at Focus (mm)

< 0.010

Limits for 19-ID


Parameter

Limits

Photon Energy (keV)

 

lower limit

6.0

upper limit (Si 111 crystal)

19.5

Energy Bandwidth (eV)

 

nonresonant crystallography @ 10.4 keV

0.013

resonant @ 6.5 keV

1.34

resonant @ 20.0 keV

0.84

Harmonic Contamination (%)

0.036%

Minimum Focus Achieved (mm)

 

vertical

0.038

horizontal

0.083

Maximum Accepted Divergence from Source (mrad)

 

vertical @ 12.0 keV

0.04

horizontal @ 12.0 keV

0.08

Beam Position Stability at Focus (mm)

< 0.010

Maximum Specimen to Detector Distance with A-frame (mm)

1200

ra/nd

 